CARATTERISTICHE E SENSIBILITA' DI VARIE SONDE E TUBI GEIGER MULLER

Model GP-200 End-Window G-M Probe

The Model GP-200 End Window G-M probe is sensitive to most Alpha, Beta & Gamma Radiation. This end window probe is useful for general purpose applications.

RANGE: 0-200 mR/hr Dose (based on Cs 137 calibration) Requires Linearization over 100 mR/hr
ENERGY REPONSE: Alpha: 3.5MeV minimum, Beta: 100 keV mininum, Gamma: 20 keV— 2MeV
OPERATING VOLTAGE: 900 vdc
WINDOW DENSITY: 1.6 mg/cm2 Window Density
DEAD TIME: 150 μSec
SENSITIVITY TO Cs 137: 2,500 CPM/mR/hr
SENSITIVE AREA: 9 cm2
ENVIROMENTAL: -20°F (-28° C)— 140°F (60° C) — 5 - 95% RH
IPX4 HOUSING
DIMENSIONS: 1.2” (3cm) DIAMETER X 5.5” (10cm) LONG
WEIGHT: 0.5 lbs(.23 kilos)


CBM20 / SBM-20, SBM-20U - HARD BETA and GAMMA GM DETECTORS

GENERAL SPECIFICATIONS
Parameters SBM-20 / SBM-20U
Gas Filling Ne + Br2 + Ar
Cathode Material Stainless Steel, 50 mkm
Maximum Length (mm) 108 / 101
Effective Length (mm) 91.0 / 83.5
Maximum Diameter (mm) 11
Effective Diameter (mm) 10
Connector Pin
Operating Temperature Range 0C -60 to +70
WALL SPECIFICATIONS
Parameters SBM-20 / SBM-20U
Areal Density (mg/cm2) 40
Thickness (mm) 0.05
ELECTRICAL SPECIFICATIONS
Parameters SBM-20 / SBM-20U
Minimum Anode Resistor (meg ohm) 1.0
Recommended Anode Resistor (meg ohm) circuit diagram 5.1
Recommended Operating Voltage (volts) 400
Operating Voltage Range (volts) 350 - 475
Initial voltage (volts) 260 - 320
Plateau length (volts) at least 100
Maximum Plateau Slope (%/100 volts) 10
Minimum Dead Time (at U=400V, micro sec) 190
Working range (mkR/s) 0.004 - 40
Working range (mR/h) 0.014 - 144
Gamma Sensitivity Ra226 (cps/mR/hr) 29
Gamma Sensitivity Co60 (cps/mR/hr) 22
Inherent counter background (cps) 1
Tube Capacitance (pf) 4.2
Life (pulses) at least 2*1010
Weight (grams) 10 / 9

TUBO PANCAKE RUSSO #2

Besides beta-particles, the counter can register alpha-particles and gamma radiation. To register alpha-and beta-particles the counter #2 is provided with a thin mica window.
The other design peculiarity is a minimal possible cathode surface (in comparison with cylindrical end-window beta-counters) that allows registering alpha-and beta-radiation

MAIN DATA:
Work. Voltage, V 400
Plateau Length/Inclination, V / %/V 200 / 0.25
Own Background, Pulses/s 1
Sensitiv. to Gamma Radiation 60Co or 137Cs, not less, Pulses/uR 160
Sensitiv. to Gamma Radiation 60Co or 137Cs, not more, Pulses/uR 240
Efficiency to Beta-Radiation 90Sr-90Y, not less, % 60
Efficiency to Beta-Radiation 204Tl, not less, % 45
Efficiency to Alpha-Radiation 239Pu,not less, % 20
Window Area, cm2 13.8-15.2
Mica Thickness, ?m 10-14
Operating temprerature range, 0C +60...-50
Service Life , Pulses 1x1010
Weight, gram 27.5